Transmitter-receiver coil separation
is one of many factors affecting the depth of penetration
and background noise levels for Helicopter ElectroMagnetic
(HEM) surveys. Some of the others are transmitter power,
bird stability, electronic noise levels and real-time
filter quality.
Frequency domain HEM systems transmit
a primary field, and measure the secondary response
from the earth in the presence of the primary. As a
consequence, the response is generally measured normalised
to the strength of the primary field, in parts per million.
The minimum measurable signal is defined by the smallness
of signal that can be resolved in the presence of the
very large primary field and is usually defined in ppms.
Therefore, the ppms of one system will differ from another.
The depth of detection and the fineness
of resistivity resolution, depend on the ability to
resolve weak signals and subtle changes in signal in
the presence of the noise. "Noise" includes
both high-frequency variations in the range of anomaly
wavelengths, and drift, which interferes with the system's
ability to map resistivity accurately over a large area.
The smaller the actual ppm value is,
for the same ppm noise level, the finer the system resolution.
This creates an apparent contradiction: the more powerful
the transmitted field, the better, but the smaller the
primary field interfering with the measurement at the
receiver, also the better. This is best accomplished
by having the receiver as far as practical from a powerful
transmitter. The farther away the receiver is from the
transmitter, the lower the primary field at the receiver
will be, allowing for the detection of smaller secondary
fields. (When the system altitude is significantly higher
than the transmitter-receiver separation, the response
from the earth depends only on the transmitter power
and frequency.)
The decrease in primary field is a
function of the coil separation cubed. Thus for an increase
in coil separation from 6m to 8m, the decrease in received
primary field would be (8/6)3=2.4 times. The anomaly
responses, measured in ppms, would be 140% bigger.
Most systems measure their noise level
in ppm, and about 2 ppm is common. To compare systems,
the noise level of each must be divided by the coil
separation cubed. To compare two systems with claimed
noise levels of 2 ppm, one with 5 m coil separation
and one with 8 m coil separation, the relative lower
noise limits would be 2/53=0.016 to 2/83=0.004 for 5
m and 8 m respectively. This means that the 8 m bird
ppms have a resolution 4 times better than the 5 m bird.
The closest analogy to this effect
which is familiar to most geologists is ground HLEM
, such as the Max-Min. The Max-Min uses longer coil
spacing (up to 250m) to see deeper, because the receiver
is farther from the transmitter, so the secondary field
is much larger relative to the primary. (The absolute
secondary field is actually weaker, since the transmitter
and receiver are farther from the conductor.) Unlike
Max-Min, because the coils are much farther above the
ground than they are from each other, the coil separation
of a HEM system does not effect spatial resolution.
Greg Hodges, Chief Geophysicist, 2001